Nasopharyngeal airway size selection and its implication in the management of pediatric difficult airway


Department of Anaesthesiology, Jawaharlal Institute of Post Graduate Medical Education and Research (JIPMER), Puducherry, India

Date of Submission10-Jun-2019
Date of Acceptance28-Oct-2019
Date of Web Publication18-Jan-2021

Correspondence Address:
Dr. Sakthirajan Panneerselvam
27&28, 1st Main Street, Priyadarshini Nagar, Puducherry - 605 006
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/joacp.JOACP_187_19

Rights and Permissions

How to cite this article:
Thangavel AR, Panneerselvam S, Rudingwa P, Sivakumar RK. Nasopharyngeal airway size selection and its implication in the management of pediatric difficult airway. J Anaesthesiol Clin Pharmacol 2020;36:565-6

How to cite this URL:
Thangavel AR, Panneerselvam S, Rudingwa P, Sivakumar RK. Nasopharyngeal airway size selection and its implication in the management of pediatric difficult airway. J Anaesthesiol Clin Pharmacol [serial online] 2020 [cited 2021 Mar 4];36:565-6. Available from: https://www.joacp.org/text.asp?2020/36/4/565/307192



Dear Editor,

Choosing an appropriate size nasopharyngeal airway (NPA) is essential to maintain airway patency and optimize ventilation while minimizing complications. We report our experience with the NPA while attempting to intubate a 13-year-old boy weighing 25 Kg, who had unilateral congenital temporomandibular joint ankylosis with micrognathia and was scheduled for mandibular distraction surgery. We had planned to perform fiberoptic bronchoscope (FOB) guided intubation while maintaining spontaneous ventilation following an inhalational induction. Under standard monitoring, anesthesia was induced with incremental doses of sevoflurane in oxygen. With increasing anesthetic depth, the child developed substernal retractions and paradoxical abdominal movements indicating respiratory distress which were not relieved with head tilt-chin lift maneuver. Hence, a well-lubricated 6 mm internal diameter NPA (Argyle™, Covidien, Ireland) was inserted to relieve the obstruction. The NPA size was selected based on the nares to tragus distance and nostril diameter of the child. In spite of multiple adjustments and advancing the NPA till its flange, airway obstruction persisted. Suspecting that the previously inserted NPA was insufficient in length to relieve the airway obstruction, we replaced it with a second NPA (Rusch, Teleflex, USA) of same internal diameter and the length of which was a little longer than the former (130 mm vs. 120 mm, [Figure 1]). This instantly relieved the airway obstruction, as confirmed by relaxed breathing pattern and regular capnographic trace. Subsequently FOB-guided intubation was successfully performed through the other nostril and the tip of the NPA was observed to be just in front of the larynx effectively bypassing the tongue and epiglottis.
Figure 1: (a and b) Two NPA from different manufacturers of same internal diameter but different lengths

Click here to view


The investigation has suggested that the external diameter of nares and little finger did not aid in appropriate selection of NPA in a magnetic resonance image-based study.[1] Stoneham found that clinical measurements like nares to tragus distance, nares to mandible distance and mento-hyoid distance did not correlate with the nares to epiglottis (NE) and larynx (NL) distance. It was found that the NE and NL distances had a wide range and correlated best with height.[2]

Although pediatric nomograms to select appropriate NPA based on nares to mandible distance, and height of the patient are available,[3] their use may be limited by variations in anatomy. The selection of ideal length and diameter NPA beforehand is important to avoid the risk of repeated manipulations and hence airway trauma and bleeding. In a situation like ours, where FOB-guided intubation was planned, it was essential to have a bloodless field. It is clear from the literature that the length more than the diameter is important for successful NPA placement and the NE distance is a highly variable parameter even within the same age group. Commercially available NPA have different length for a given internal diameter, which varies among manufacturers. We suggest choosing an NPA based on the diameter that has a maximum length preferably with an adjustable flange. In a spontaneously breathing patient, advance the NPA gently till the air blast is felt or capnographic trace is obtained, and fix it at that length. Ideal position of the tip of NPA will be just above the epiglottis, so that airway obstruction is relieved. And If inserted beyond, it may enter the larynx causing cough, laryngospasm, and trauma to cords or enter into esophagus causing gastric insufflation, nausea, and vomiting.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form, the patient has given his consent for his images and other clinical information to be reported in the journal. The patient understands that his name and initials will not be published and due efforts will be made to conceal his identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Roberts K, Allison KP, Porter KM. A review of emergency equipment carried and procedures performed by UK front line paramedics. Resuscitation 2003;58:153-8.  Back to cited text no. 1
    
2.
Stoneham MD. The nasopharyngeal airway. Assessment of position by fibreoptic laryngoscopy. Anaesthesia 1993;48:575-80.  Back to cited text no. 2
    
3.
Kim SH, Kim DH, Kang H, Park JJ, Seong SH, Suk EH, et al. Estimation of the nares-to-epiglottis distance and the nares-to-vocal cords distance in young children. Br J Anaesth 2012;109:816-20.  Back to cited text no. 3
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
   References
   Article Figures

 Article Access Statistics
    Viewed172    
    Printed2    
    Emailed0    
    PDF Downloaded24    
    Comments [Add]    

Recommend this journal